:DISK:SAVE:IMAGe

Command
:DISK:SAVE:IMAGe "<file_name>" [,<format>
[,{SCReen | GRATicule}
[,{ON | 1} | {OFF | 0}
[,{NORMal | INVert}
[,{ON | 1} | {OFF | 0}]]]]]
The DISK:SAVE:IMAGe command saves a screen image. The default path is
C:\Document and Settings\All Users\Shared Documents\Infiniium\Data.
<format>
The image format can be: BMP, GIF, TIF, PNG, or JPEG. The extension is
supplied by the oscilloscope depending on the selected file format.
If you do not include the format in the command, the file is saved in the
format shown in the Save Screen dialog box.
<file_name>
A quoted ASCII string with a maximum of 254 characters including the
entire path name, if used.
(First) ON | OFF
ON means that compression is on for the bitmap format (BMP). OFF
means compression is off.
(Second) ON |
The second ON/OFF selection indicates to save the setup information in
OFF
the image or not.
<format>
{BMP | GIF | TIF | JPEG | PNG}
Examples
myScope.WriteString ":DISK:SAVE:IMAGE ""FILE1"",BMP,SCR,ON,INVERT"
or:
myScope.WriteString ":DISK:SAVE:IMAGE ""FILE1"",TIF,GRAT,ON"
or:
myScope.WriteString ":DISK:SAVE:IMAGE ""FILE1"""