6
Status Reporting

Trigger Event Register

This register sets the TRG bit in the status byte register when a trigger
event occurs.
The trigger event register stays set until it is cleared by reading the
register with the TER? query or by using the *CLS (clear status)
command. If your application needs to detect multiple triggers, the trigger
event register must be cleared after each one.
If you are using the Service Request to interrupt a computer operation
when the trigger bit is set, you must clear the event register after each
time it is set.