6500 Series

6510-
Main pressure transducer overrange detected.
6511-
Sample transducer (number) overrange detected.
6512-
Po transducer (number) overrange detected.
6513-
Hardware pressure overrange detected.
Cause:
Action:
6516-
Sample pressure greater than (mmHg) is not allowed.
Cause:
Action:
6517-
(Unit [number]) Total volume dosed greater than (volume) cm3 is not allowed.
Cause A:
Action A:
Cause B:
Action B:
Cause C:
246-42800-01 - Dec 2013
A manifold pressure over 1000 mmHg was detected by the hardware.
Observe caution when operating the analyzer manually. If the problem
persists, contact your service representative.
An absolute pressure greater than (pressure) mmHg was attained that
exceeded the specified maximum manifold pressure.
The analysis was canceled. All previously collected data were stored.
3
More than (volume) cm
to leaks in the system.
Perform leak checks on the system. Refer to Appendix D,
THE ANALYZER FOR
The value for the Po may be too low, causing the instrument to mistake
condensation for actual adsorption.
Change the analysis variables to directly measure the value of Po dur-
ing the experiment.
The amount of sample you used may be too large. Multiply the max-
imum value from the Volume Adsorbed column of the Isotherm Report
by the sample weight. If the result exceeds (volume) cm
of sample is too large.
has been dosed onto the sample, possibly due
LEAKS, page D-1.
Appendix B
TESTING
3
, the amount
B-41