1
Performance Verification Tests

Oscilloscope Performance Verification Tests

The performance verification tests described in this section verify the channel
measurement accuracy and acquisition system to provide confidence in
functionality and accuracy of the U1610/20A. These performance verification
tests are based on the U1610/20A specifications, as listed in the U1610/20A
User's Guide.

Voltage measurement accuracy verification test

This test verifies the accuracy of the analog channel voltage measurement for
each channel (DC vertical gain accuracy and dual cursor accuracy specifications).
In this test, you will measure the DC voltage output of an oscilloscope calibrator
using dual cursors on the U1610/20A.
Test limit:
±{4% of full scale* + 0.4% of full scale* (~1 LSB)}
*Full scale on all ranges is defined as 8 divisions multiplied by the V/div setting.
Refer to
accuracy verification test. Perform the following procedure for channel 1 and then
repeat the same procedure for channel 2.
1 On the U1610/20A:
a Press
b Press
Offset adjustment is not applicable for the 50 V/div vertical sensitivity.
NOTE
20
All manuals and user guides at all-guides.com
Table 1-1
for the equipment required for this voltage measurement
>
(F2) and use the
offset to 0.5 major divisions from the bottom of display.
>
(F1) to turn on channel 1.
or
key, or press
Keysight U1610/20A Service Guide
to set the