2 Connect the probe's retractable hook tip to the point of interest on the circuit or
CAUTION
135 Vrms
When measuring voltages over 30 V, use a 10:1 probe.
CAUTION
CAUTION
Defeating the ground connection and "floating" the oscilloscope chassis will probably
result in inaccurate measurements and may also cause equipment damage. The probe
ground lead is connected to the oscilloscope chassis and the ground wire in the power
cord. If you need to measure between two live points, use a differential probe with
sufficient dynamic range.
Do not negate the protective action of the ground connection to the oscilloscope. The
WARNING
oscilloscope must remain grounded through its power cord. Defeating the ground
creates an electric shock hazard.

Input a Waveform

The first signal to input to the oscilloscope is the Demo 2, Probe Comp signal. This
signal is used for compensating probes.
1 Connect an oscilloscope probe from channel 1 to the Demo 2 (Probe Comp)
2 Connect the probe's ground lead to the ground terminal (next to the Demo 2
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
device under test. Be sure to connect the probe ground lead to a ground point
on the circuit.

Maximum input voltage at analog inputs

Do not float the oscilloscope chassis

terminal on the front panel.
terminal).
1
Getting Started
29