ADJUSTMENTS
5
.
Open the INPUT FLAP.
Select OPEN
6
.
Adjust SENSOR B3/C_IF_EO INPUT FLAP OPEN that the FLAP is fully open.
7
.
Exit the SERVICE MODE.
Press 3 times BACKSPACE
Select LEAVE COMPONENT TEST ................................................................... press ENTER
Select QUIT ML300 SERVICE MODE ............................................................... press ENTER
Select Quit the program ................................................................................. press ENTER.
CASSETTE REGISTRATION SENSOR B2.
Purpose:
To make sure that the SENSOR B2 is not deactuated by a curved CASSETTE before it is
removed from the ML300.
1
.
Take out the MAGAZINES to avoid FILM FOGGING.
2
.
Take off the PANELS.
3
.
Check that the distance between the CASSETTE REGISTRATION ROLLER level
and the TRANSPORT ROLLER level is ≥ 3mm.
KODAK AG, Stuttgart
figure 3-4
3-5
SM 3058-6
9/98