ADJUSTMENTS
85.
Exit the SERVICE PROGRAM.
Press 3 times BACKSPACE
Select LEAVE COMPONENT TEST ................................................................... press ENTER
Select QUIT ML300 SERVICE MODE ............................................................... press ENTER
Select Quit the program ................................................................................. press ENTER
86.
Put the FILM CHUTE in its correct position and fix it with all 4 SCREWS.
87.
Run several cycles with X-OMAT CASSETTES and with VIDEO-FILM HOLDERS
(if available).
88.
Make sure that the MAGAZINE BLOWPIPES do not interfere with the CASSETTES.
Note
This adjustment is most important when OPERATING SOFTWARE VERSION 3.0 AND VIDEO
FILM HOLDERS are used, because in this combination the parameter "LOWER FILM
POCKET" is ignored. If it is not correct, the FILM may float out of the VHF.
KODAK AG, Stuttgart
figure 3-55
3-57
SM 3058-6