ADJUSTMENTS
10.
Check if the voltage reading is ≤ 1.0V. If the voltage is correct proceed with step
13, else proceed with step 11.
11.
Adjust the position of SENSOR B22 until the voltage reading reaches the minimum.
12.
Adjust R101 on PCB A8 until the voltage reaches a minimum of ≤ 1.0V. Do not
adjust R101 any further.
13.
Check that the SENSOR B22 is indicated as off when the reflective sticker is
covered. The voltage between TP14 and TP 10 should be > 4.0V. If X-Omatic
CASSETTES (yellow) are used, insert a CASSETTE without reflective stickers and
check if the voltage is > 3.5V. If the voltage is correct proceed with step 14. If the
voltage is not correct, increase it with R101 on PCB A8 until it reaches >3.5 V. Now
feed in a CASSETTE with reflective stickers and repeat step 10 to 13.
14.
Exit the SERVICE MODE.
Press 3 times BACKSPACE .
Select LEAVE COMPONENT TEST ................................................................... press ENTER
Select QUIT ML300 SERVICE MODE ............................................................... press ENTER
Select Quit the program ................................................................................. press ENTER
15.
Take out the CASSETTE.
16.
Mount the PANELS.
17.
Check with different TYPE 2 CASSETTES (MAMMO and X-OMAT) for correct
operation.

CASSETTE LENGTH

Purpose:
This adjustment ensures that the correct CASSETTE LENGTH is measured.
Warning
BE CAREFUL WHEN WORKING IN THE CASSETTE OPENER AREA. THE OPENER MOTOR
AND THE OPENER MECHANISM ARE VERY STRONG. THEY CANNOT BE MOVED
MANUALLY. THEY MAY SQUEEZE YOUR HAND AND TRAP YOU IF YOU TRY TO STOP
THEM MANUALLY. NEVER START THE CASSETTE OPENER MOTOR WHEN SOMEONE'S
HANDS ARE IN THE CASSETTE AREA.
KODAK AG, Stuttgart
3-25
SM 3058-6
9/98