SM 3058-6
13.
Switch on the XML 300.
Warning
Keep your hands out of the CASSETTE OPENER AREA when doing these measurements.
14.
This adjustment is sensitive to vibrations of the DRIVE SHAFT. Therefore it is not
enough to do it in the static mode. It has to be repeated in the dynamic mode as well.
PART 2:
This procedure is valid for ML300 with SN 1430 and up.
!
Caution
Take proper ESD SAFETY PRECAUTIONS when doing this adjustment.
1
.
Start the SENSOR TEST.
Start the SERVICE PROGRAM
Select SERVICE MODE from the GLOBAL MENU ....................................... press ENTER
ENTER SERVICE MODE MESSAGE is displayed............................................ press ENTER
UNIT DATA are displayed ................................................................................ press ENTER
Select COMPONENT TEST from the MAIN MENU ....................................... press ENTER
Select SENSORS .................................................................................................. press ENTER
Select SENSOR TEST WITH SOUND.................................................................. press ENTER
2
.
Set the JUMPER of A8X50 to the upper position (pin 3-4).See the drawing on the
next page.
3
.
Adjust the position of the MIRROR of B20, so that B20 is indicated as off.
Use all types of CUSTOMER FILM to ensure correct operation.
SENSOR B20
MIRROR
9/98
figure 3-30
3-32
ADJUSTMENTS
KODAK AG, Stuttgart