Page 2
Eastman Kodak Company reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information. Kodak shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by Kodak’s negligence or other fault.
Page 3
Weekly Preventive Maintenance Section 1: Weekly Preventive Maintenance Important The operator of the PROCESSOR should do the weekly preventive maintenance procedures in this section. WET SECTION COVER 2 EVAPORATION COVERS DEVELOPER FILTER WATER RESERVOIR ENTRANCE CROSSOVER DEVELOPER RACK DEVELOPER/FIXER WATER DISCONNECT CROSSOVER FIXER/ FIXER...
Page 4
PERIODIC PREVENTIVE MAINTENANCE [9] Check that the CROSSOVER TROUGHS are clean, are not cracked or broken. [10] Wipe any deposits from the processing section of the PROCESSOR. [11] Install the following parts. Check that each assembly seats correctly: • WASH RACK •...
Page 5
Monthly Preventive Maintenance Section 2: Monthly Preventive Maintenance Important Completion of the monthly preventive maintenance procedures is recommended to ensure the optimum reliability of the PROCESSOR. Warning Dangerous Voltage. [1] De-energize the PROCESSOR. [2] Remove: • WET SECTION COVER • WATER DISCONNECT and WATER RESERVOIR •...
Page 6
PERIODIC PREVENTIVE MAINTENANCE [11] Check that the SLOTS in the AIR TUBES are clean. If the parts are dirty: (a) Remove the BAFFLES from the AIR TUBES. (b) Rinse the BAFFLES with water. (c) Clean the AIR TUBES with a BOTTLE BRUSH and rinse with water. [12] Wipe any deposits from the processing section of the PROCESSOR.
Page 7
[7] Clean the FIXER RACK, the WASH RACK, and the FIXER TANK with Kodak FIXER/WASH SYSTEM CLEANER, CATALOG No. 139 5110. [8] Clean the WASH TANK with Kodak FIXER/WASH SYSTEM CLEANER or, if biological growth exists, with a 10% solution of chlorine bleach.
Page 8
PERIODIC PREVENTIVE MAINTENANCE Important Do not use abrasive cleaning materials on the ROLLERS. [10] Rinse the following parts with warm water, 44C (110F) or less. Wipe the ROLLERS and the GUIDE SHOES with a soft, synthetic sponge: • CROSSOVERS • EVAPORATION COVERS •...
Page 9
3 Month Preventive Maintenance Caution • Any obstruction such as biological growth, tight bends, or an upward sloping WASH DRAIN HOSE can cause wash water to return to the WASH TANK causing an overflow of water and damage to the ELECTRICAL BOX. •...
Page 10
[31] Check the negative pressure at the EXHAUST VENT for the DRYER. See the MULTILOADER 300 PLUS INSTALLATION INSTRUCTIONS. [32] Calibrate the REPLENISHMENT PUMPS. See the section “Calibrating the Replenishment System” in the OPERATOR'S MANUAL for the Kodak X-Omat MULTILOADER 300 PLUS. 28JAN98 – 7C8289...
Page 11
= Apply recommended lubricant. C(1) = Clean with Kodak DEVELOPER SYSTEM = Replace, if necessary. CLEANER AND NEUTRALIZER. = Test and adjust, if necessary. C(2) = Clean with Kodak FIXER/WASH SYSTEM CLEANER.
Page 12
Kodak and X-Omat are trademarks. Printed in U.S.A. • pm3488_1.fm EASTMAN KODAK COMPANY HEALTH IMAGING Rochester, NY 14650...