ADJUSTMENTS
3
.
Connect PIN 1 to PIN 3 at the SOCKET of U41.
4
.
Connect the DVM to TP12 of PCB A4 and to TP11(GND PE) of PCB A4 (for the
location of the TEST PINS see the drawing on the next page ).
5
.
Switch on the XML300.
6
.
Adjust the mechanical position of B35 relative to its MIRROR, so that the indicated
voltage is < 500mV.
KODAK AG, Stuttgart
figure 3-78
SENSOR B35
figure 3-79
3-89
SM 3058-6
9/98