SM 3058-6

HUMIDIFIER (MOD 14)

1
.
To reduce the potential for biological growth and contamination within the ML300
HUMIDIFIER, the water reservoir should be drained and the cleaned with a disinfectant.
The FILTER 9198261 should be replaced at least every 6 month. The biological growth
and contamination are not originated by the HUMIDIFIER. They are caused by the
outside air and the water. They are then accumulated in the FILTER. The disinfectant
has to be bought locally.
Warning
Follow the how-to-use and the safety instructions supplied with the disinfectant.
The DRAIN must never be sealed
2
.
To avoid flooding of the ML300 the DRAIN HOSE must be connected to an open
DRAIN The safety opening of the Y-CONNECTOR must not be clogged.
9/98
FILTER 9198261 is below this COVER
EXTERNAL VERSION (MOD 14)
figure 10-4
10-6
PREVENTIVE MAINTENANCE
The open end of the Y-CONNECTOR must
point to the right. It must never be closed.