ADJUSTMENTS
Select CHANGE PARAMETER .......................................................................... press ENTER
Select CASSETTE LENGTH ................................................................................. press ENTER
Note
A small CASSETTE preferably 18x24 should be used.
14.
Key in the measured length (NOT THE FILM SIZE !!!) of the selected CASSETTE
in mm.
15.
Insert the selected CASSETTE and press ENTER. The ML300 measures the
CASSETTE LENGTH and calculates the correction value.
16.
This correction value must be stored in the ML300 MEMORY.
Select STORE PARAMETERS ............................................................................. press ENTER
17.
Take out the CASSETTE.
18.
Exit the SERVICE PROGRAM.
Press BACKSPACE twice
Select QUIT ML300 SERVICE MODE ............................................................... press ENTER
Select Quit the program ................................................................................. press ENTER
19.
Close the TOP COVER.

SENSOR B20 VACUUM OFF

Purpose:
This adjustment ensures that the LEADING EDGE of the exposed CASSETTE FILM is
detected to turn off the CASSETTE SUCKER BAR VACUUM. Use only SENSORS
P/N 9228991. They are selected and known to be good.
Note
·
Part 1 is valid for ML300 up to SERIAL NUMBER 1429. Part 2 is valid for ML300 with
SERIAL NUMBER >01430.
·
Take proper ESD SAFETY PRECAUTIONS when doing this adjustment.
PART 1:
The procedure is valid for all XML 300, no matter if CONNECTOR A8x50 has 3 pins or 6
pins. If it has 6 pins, leave the jumper in the position where it was set by the Manufacturer. It
must be in the middle position (pin 2-5)
KODAK AG, Stuttgart