SM 3058-6
5
Check if the voltage reading is ≤ 1.0V. If the voltage is correct, proceed with step 8,
.
if not proceed with step 6.
6
.
Adjust the position of SENSOR B21 until the voltage reading reaches the minimum.
7.
Adjust R100 on PCB A8 until the voltage reaches a minimum voltage of ≤ 1.0V. Do
not adjust R100 any further.
8.
Check that the SENSOR B21 is indicated as off when the reflective sticker is
covered. The voltage between TP 15 and TP 10 should be > 4.0V. If X-Omatic
CASSETTES (yellow) are used, insert a CASSETTE without reflective stickers and
check if the voltage is > 3.5V. If the voltage is correct proceed with step 9. If the
voltage is not correct, increase it with R100 on PCB A8 until it reaches >3.5 V. Now
feed in a CASSETTE with reflective stickers and repeat step 5 to 8.
9.
Connect the DVM between TP14(SENSOR B22) and TP10 GND PE on PCB A8.
9/98
figure 3-24
3-24
ADJUSTMENTS