ADJUSTMENTS
4
.
Connect the DVM to TP11 on PCB A8 and to TP10 (GND PE) on PCB A8. See the
figure on the next page.
Warning
You are now working in the CASSETTE OPENER AREA. Be very careful that the CASSETTE
OPENER is not started by accident. Do not override the INTERLOCK SYSTEM.
5
.
Switch on the XML 300.
6
.
Adjust the mechanical position of SENSOR B20 relative to its MIRROR, so that the
indicated voltage is < 500mV.
SENSOR B20
MIRROR
KODAK AG, Stuttgart
figure 3-26
figure 3-27
3-29
SM 3058-6
9/98