ADJUSTMENTS
4
Adjust R38 to 1.88V ±50mV. This results in a phase current of 4A.
.
5
.
Disconnect the DVM.
6
.
Exit the SERVICE MODE.
Press 3 times BACKSPACE
Select LEAVE COMPONENT TEST ................................................................... press ENTER
Select QUIT ML300 SERVICE MODE ............................................................... press ENTER
Select Quit the program ................................................................................. press ENTER
KODAK AG, Stuttgart
figure 3-67
3-77
SM 3058-6
9/98