Eastman Kodak Company reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information. Kodak shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by Kodak’s negligence or other fault.
Weekly Preventive Maintenance Section 1: Weekly Preventive Maintenance Important The operator of the PROCESSOR should perform the preventive maintenance procedures in this section weekly. DEVELOPER/ FIXER EVAPORATION CROSSOVER COVERS FIXER/WASH CROSSOVER SQUEEGEE ASSEMBLY DETECTOR CROSSOVER DEVELOPER RACK CROSSOVER CLEANING FIXER YOKE RACK WASH...
Page 4
PERIODIC/PREVENTIVE MAINTENANCE Important When you install the parts in the following steps, check that each part seats firmly. [12] Install: a. WASH RACK b. all CROSSOVERS [13] Connect the QUICK DISCONNECT VALVE. YOKE FIXER/WASH DEVELOPER/ CROSSOVER FIXER CROSSOVER QUICK DISCONNECT VALVE H148_0130BCA H148_0130BA...
Monthly Preventive Maintenance Section 2: Monthly Preventive Maintenance Important Completion of the monthly preventive maintenance procedures is recommended to ensure the optimum reliability of the PROCESSOR. [1] De-energize the PROCESSOR. [2] Remove the TOP COVER. [3] Remove the following parts: a.
Page 6
PERIODIC/PREVENTIVE MAINTENANCE [15] Allow the DETECTOR CROSSOVER to air dry before you install it in the PROCESSOR. [16] Clean the YOKE. Check that the holes in the YOKE are free of biological growth and other obstructions. [17] Install the RACKS. Important When you install the parts in the following steps, check that each part seats firmly.
[7] Clean the DEVELOPER RACK and the DEVELOPER TANK with Kodak DEVELOPER SYSTEM CLEANER AND NEUTRALIZER, CAT No. 843 4615. [8] Clean the FIXER RACK, the WASH RACK, and the FIXER TANK with Kodak FIXER/WASH SYSTEM CLEANER, CAT No. 139 5110.
Page 8
PERIODIC/PREVENTIVE MAINTENANCE [9] Clean the WASH TANK with Kodak FIXER/WASH SYSTEM CLEANER or, if biological growth exists, with a 10% solution of chlorine bleach. [10] If necessary, clean the LEVEL SENSOR PROBES with the correct SYSTEM CLEANER. [11] Rinse the following parts with warm water, 44[degree]C (110[degree]F) or less. Wipe the ROLLERS and the GUIDE SHOES with a soft, synthetic sponge.
Page 9
3-Month Preventive Maintenance [25] Check the water supply FILTER. If necessary, change the FILTER. Caution • Any obstruction such as biological growth, small bends, or an upward sloping WASH DRAIN HOSE can cause wash water to return to the WASH TANK causing an overflow of water and damage to the ELECTRICAL BOX or COMPONENT PLATE.
† Completion of the monthly preventive ‡ The 3-month schedule is a guideline for sites maintenance procedures is recommended to adhering to the site specifications issued by Kodak. ensure the optimum reliability of the Actual usage and site characteristics may PROCESSOR.
3-Month Preventive Maintenance Section 5: Publication History Print Date Pub. No. ECO No. Affected Pages File Name Notes October 1995 5B6747 2649-029 pm3425_1_029.doc First printing. January 1996 5B6747 2649-062 Covers pm3425_1_062.doc Graphic Unification. January 1998 5B6747 2649-062 pm342500.fm First CD-ROM printing. Content is identical to January 1996 version;...
Page 12
Kodak and X-Omat are trademarks. HEALTH IMAGING pm342500.fm EASTMAN KODAK COMPANY ROCHESTER, N.Y. 14650 Printed In USA...