Venting - Kodak X-Omat MULTILOADER 300 Plus Service Manual

Hide thumbs Also See for X-Omat MULTILOADER 300 Plus:
Table of Contents

Advertisement

SM 3058-6

VENTING

1
.
Check that the ML300 is vented correctly. Use the following procedures to check the
negative air flow of the site.
Note
This procedure is designed to check the negative exhaust flow of the building not the
processor.
2
.
Verify that the processor is switched off.
3
.
Use AIR METER TL-2431 to measure negative static pressure in the EXHAUST
DUCT 30.5 cm (12 in.) from the end that is connected to the processor.
9/98
AIR METER TL2431
figure 10-6
10-8
PREVENTIVE MAINTENANCE
KODAK AG, Stuttgart

Advertisement

Table of Contents
loading

This manual is also suitable for:

X-omat multiloader 300

Table of Contents