5.32
IP Removal Arms
{SP:INDEX}
u NOTE u
The replacement procedures for the IP removal arms for the first, second, third, and
fourth shelves are all identical.
Removal Procedures
1.
Move the arms (white) to the positions illustrated below.
2.
Remove the arms (white).
{MC:5.6_Arms (White)}
011-202-08E
09.30.2009 FM5540
3.
Remove the suction pump bracket assembly.
INSTRUCTION
u
u
Remove the suction pump bracket assembly for the shelf where the IP removal
arms to be removed is installed.
{MC:5.31_Suction Pump Bracket Assemblies}
4.
Remove the bearings.
{CHECK 1}
CR-IR 363 Service Manual
MC-65