4.
Check the CL's IP address and record it.
u NOTE u
The CL's IP address must be restored to the user specified value after the RU
application is initialized. Be sure to confirm the CL's IP address (user specified
value) and record it.
5.
Change the CL's IP address to "172.16.1.20".
6.
Turn ON the power of the RU.
The RU boots up to the following state.
- The touch panel does not display anything and remains blank.
- Only the cassette IN and OUT lamps are illuminated.
011-202-04
11.15.2004 FM4425
7.
Put the RU-Appl CD-ROM into the CD-ROM drive of the CL.
8.
Perform the following procedures.
u NOTE u
Always input "172.16.1.10" in RU IP ADDR.
9.
Check to ensure that the RU IP address input in the step 8 (172.16.1.10)
is displayed, and click on [OK].
CR-IR 363 Service Manual
Appx MT-2
Appx MT-2