Advertisement

3.
Perform the sensitivity correction by the use of MUTL of the
maintenance mode on the CR console.
INSTRUCTION
u
u
Enter the value (x.xx) of the X-ray dose measured at step 1 into the field of
sensitivity adjustment.
REFERENCE
If the measure value of the X-ray dose is 18.1 mR: The value to be entered
is "18.1". If the measured value of the X-ray dose is 20 mR: The value to be
entered is "20".
011-202-11E
06.29.2012 FM6046
4.
Read the IP exposed in step 2 for a one-minute period from 10 to 11
minutes after the exposure by the use of "TEST", "image format" or
"image format-2" and "SINGLE" in the CR console menu, and output
the image.
INSTRUCTION
u
u
Be sure to follow the instruction mentioned in "20.2 Image/Conveyance Checks"
when outputting the image.
{MC:20.2_Image/Conveyance Checks}
5.
Make sure that "RESULT-OK" appears on screen.
6.
Using the X-ray dosimeter, measure the X-ray dose for exposing the
IP, and write it down.
7.
Expose an IP of 18"x 24" HR-BD or 24"x 30" HR-BD size with a
molybdenum X-ray tube (approx. 20 mR).
8.
Read the IP exposed in step 7 for a one-minute period from 10 to 11
minutes after the exposure by the use of "TEST", "image format" or
"image format-2" and "SINGLE" in the CR console menu, and output
the image.
Make sure that the S value is normal with the measured X-ray dose and that
there is no particular irregularity.
{MC:19.2_Checking the Image}
REFERENCE
For 2 mR: The S value is 1200.
For 20 mR: The S value is 120.
For 200 mR: The S value is 12.
9.
Back up the data so adjusted into a floppy diskette.
{MC:19.1_Backing Up the Scanner Data}
CR-IR 363 Service Manual
MC-420
MC-420