FujiFilm PROFECT CS Service Manual page 316

Table of Contents

Advertisement

10.6.1
Check Procedures via Monitoring Function (3)
1.
Start the RU PC-TOOL.
{MU:4.2.1_Starting RU PC-TOOL}
2.
Start monitoring of the RU PC-TOOL.
{MT:10.2.1_Starting the Monitoring Function}
3.
Select IP Removal Unit.
4.
Execute Suction Cup Arm HP corresponding to the sensor to be
checked.
l To check SA4: Click on the [Shelf 1] button.
l To check SA10: Click on the [Shelf 2] button.
l To check SA16: Click on the [Shelf 3] button.
l To check SA22: Click on the [Shelf 4] button.
011-202-04
11.15.2004 FM4425
5.
Check the monitoring display.
Check the display of the monitoring window of the CL.
l Example of GOOD indication
l NO GOOD indication
For the sensor to be checked, neither "OPEN" nor "CLOSE" is displayed.
u NOTE u
Click on the [START] button in the monitoring window to initiate the monitoring
beforehand. Unless the monitoring is initiated, nothing is displayed in the window
even when the sensor changes its status.
6.
Exit the monitoring of the RU PC-TOOL.
{MT:10.2.2_Exiting the Monitoring Function}
CR-IR 363 Service Manual
MT-159
MT-159