FujiFilm PROFECT CS Service Manual page 359

Table of Contents

Advertisement

10.21.1 Check Procedures via Monitoring Function (17)
1.
Start the RU PC-TOOL.
{MU:4.2.1_Starting RU PC-TOOL}
2.
Start monitoring of the RU PC-TOOL.
{MT:10.2.1_Starting the Monitoring Function}
3.
Manually raise/lower the movable guide, and check the monitoring
display.
- To check SA6: Raise/lower the movable guide of shelf 1.
- To check SA12: Raise/lower the movable guide of shelf 2.
- To check SA18: Raise/lower the movable guide of shelf 3.
- To check SA24: Raise/lower the movable guide of shelf 4.
l Example of GOOD indication
Display example when the movable guide of the shelf 1 is raised/lowered to
check the SA6.
011-202-04
11.15.2004 FM4425
l NO GOOD indication
Neither "OPEN" nor "CLOSE" is displayed for the sensor to be checked, after the
movable guide is raised/lowered.
u NOTE u
Click on the [START] button in the monitoring window to initiate the monitoring
beforehand. Unless the monitoring is initiated, nothing is displayed in the window
even when the sensor changes its status.
4.
Exit the monitoring of the RU PC-TOOL.
{MT:10.2.2_Exiting the Monitoring Function}
CR-IR 363 Service Manual
MT-202
MT-202