Error
Error Name
Code
Image data variation or image
11287
signal connector error during
bootup
11320
Cassette setting status error (1)
11321
Cassette setting status error (2)
11322
Cassette setting status error (3)
11323
Cassette setting status error (4)
11324
Hold pin release error
11325
Cassette loading inlet disabled
11328
Feed IP suction error (1)
11329
Feed IP suction error (2) (drop)
Feed/load unit entrance IP feed
11330
retry
11332
Feed IP conveyance error (1)
011-202-11E
06.29.2012 FM6046
An abnormal virtual data change was detected when the virtual data transfer diagnostics was conducted for bootup.
In the virtual data transfer diagnostics for bootup, an abnormal virtual data change, disconnected image signal
connector, or erroneous connector connection was detected.
<User operation>
After the cassette is loaded, the cassette is pulled out immediately.
<Occurrence condition>
During routine processing/M-Utility, the cassette hold solenoid (SOLA1/2/3/4) is turned OFF to hold the cassette,
and the cassette hold sensor (SA3/9/15/21) detects the OPEN state; however, the cassette IN sensor (SA2/8/14/20)
detects the OPEN state.
During routine processing/M-Utility, the cassette hold solenoid (SOLA1/2/3/4) is turned OFF to hold the cassette,
and the hold pin protrudes; however, the cassette hold sensor (SA3/9/15/21) does not transition to OPEN. Though
the retries (twice or more) are executed, the cassette is not held. (A message is notified.)
During bootup/routine processing/M-Utility, the OPEN state of the cassette IN sensor (SA2/8/14/20) is detected for
the sensor check processing just before the IP is sucked after cassette loading.
During routine processing/M-Utility, the CLOSE state of the cassette hold sensor (SA3/9/15/21) is detected for the
sensor check processing just before the IP is sucked after cassette loading.
During bootup/routine processing/M-Utility, the cassette hold solenoid (SOLA1/2/3/4) is turned ON to release the
cassette hold; however, the cassette hold sensor (SA3/9/15/21) does not transition to CLOSE. Though the retries
(twice) are executed, the cassette hold is not released. (A message is notified.)
During bootup/routine processing/M-Utility, the user disables the shelf after occurrence of "11324" error.
During routine processing/M-Utility/abnormality processing, the IP suction pump (PA1/2/3/4) is driven to convey the
IP from the cassette to the feed/load conveyor unit; however, the suction sensor (SA5/11/17/23) does not transition
to OPEN within the specified period of time. Though the retries (twice) are executed, the IP is not sucked.
During routine processing/M-Utility/abnormality processing, the IP is sucked to perform suction conveyance of
the IP from the cassette to the feed/load conveyor unit; however, the CLOSE state (IP drop) of the suction sensor
(SA5/11/17/23) is detected during IP conveyance. Though the retries (twice) for the IP suction are executed, the IP
drop is detected again.
During routine processing/M-Utility/abnormality processing, an attempt is made to convey the IP from the cassette
to the feed/load conveyor unit; however, the IP feed/load sensor (SB1/2/3/4) does not transition to CLOSE. Though
the retries (twice) for the IP suction are executed, the IP is not transported to the IP feed/load sensor.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from the
feed/load conveyor unit to the side-positioning conveyor unit; however, the pre-BCR IP sensor (SB5) does not
transition to CLOSE within the specified period of time.
Occurrence Condition
CR-IR 363 Service Manual
MT-26
Analysis
Flow
{MT:5.17}
{MT:4.21}
{MT:4.37}
{MT:4.21}
{MT:4.37}
{MT:4.37}
{MT:4.37}
{MT:4.26}
{MT:4.26}
{MT:4.22}
{MT:4.23}
MT-26