FujiFilm PROFECT CS Service Manual page 205

Table of Contents

Advertisement

Error
Error Name
Code
12822
Power supply B021 error
12823
Power supply B031 error
12824
Power supply B041 error
12825
Power supply B051 error
12826
Power supply B061 error
12827
Power supply B071 error
12828
Power supply B081 error
12829
Power supply B091 error
12830
Power supply B101 error
12831
Power supply B111 error
12832
Power supply B121 error
12833
Power supply B131 error
12834
Power supply B141 error
12835
Power supply B151 error
Power supply B series detection
12840
error (1)
12841
Power supply B error 2
011-202-11E
06.29.2012 FM6046
During bootup and during routine processing, it is detected that the "B021" fuse (F2) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B031" fuse (F3) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B041" fuse (F4) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B051" fuse (F5) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B061" fuse (F6) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B071" fuse (F7) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B081" fuse (F8) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B091" fuse (F9) on the DRV17A board is blown.
During bootup and during routine processing, it is detected that the "B101" fuse (F10) on the DRV17A board is
blown.
During bootup and during routine processing, it is detected that the "B111" fuse (F11) on the DRV17A board is
blown.
During bootup and during routine processing, it is detected that the "B121" fuse (F12) on the DRV17A board is
blown.
During bootup and during routine processing, it is detected that the "B131" fuse (F13) on the DRV17A board is
blown.
During bootup and during routine processing, it is detected that the "B141" fuse (F14) on the DRV17A board is
blown.
During bootup and during routine processing, it is detected that the "B151" fuse (F15) on the DRV17A board is
blown.
During bootup and during routine processing, even though it is detected that the voltage of the DRV17A-related
+24V power supply is not outputted, the output is detected at either of the detection points of "B011" through "B151".
During bootup and during routine processing, it is detected that the voltage of the DRV17B-related +24V power
supply is not outputted.
Occurrence Condition
CR-IR 363 Service Manual
MT-48
Analysis
Flow
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
{MT:9.}
MT-48