FujiFilm PROFECT CS Service Manual page 141

Table of Contents

Advertisement

Corresponding item
The problem is solved, that two IPs return
to the same shelf when an IP jam occurs
and the RU is rebooted.
Measures are taken against the RU
getting unreactive when the SNS17A/B
board is malfunctioning in the in-vehicle
machine.
Measures are taken against inability to
shut down when the RU is disabled.
- A function of detecting a defective lamp
temperature detection thermistor (THE1)
is added.
- Operation of warming up/cooling down
the erasure unit is improved.
The problem is solved, that the operation
panel display restores to a routine
window when the RU is rebooted and
image retransmission is completed after
the RU is disabled.
The problem is solved, that the message
of "Insufficient erasure" appears when
one of five erasure lamps is unlit during
the primary erasure.
* The RU normally works even if one
of the lamps is unlit during the primary
erasure.
The problem is solved, that the flashing
LED goes out when the IP search error
(error code: 10425) occurs during the RU
initialization.
A program is corrected, that the LED
does not light up and the alarm does not
sound when the IP position information
error (error code: 10459) occurs during
the RU initialization.
011-202-11E
06.29.2012 FM6046
RU software version
Reason
V2.0 V2.1 V2.2 V2.3 V2.4
Corrected
Corrected
Corrected
Changed
Corrected
Corrected
Corrected
Corrected
Reference
Corresponding item
When a large volume of data flows into
the network port of the RU due to wrong
network configuration or the like, the
processing has a heavy workload and the
system is disabled due to the IP overrun.
The IPL software is updated in version to
solve the problem.
A modification is made so that normal
conveyance can take place when the
cassette IN sensors (SA 2, 8, 14 and
20) are closed even if the cassette
ejection sensors (SA1, 7, 13 and 19) are
defective.
{MD:7.6}
The problem is solved, that the RU gets
unreactive and cannot boot up when
an error of "IP reading suspended:
image remains in the IP" (error code:
11499) occurs when the cassette
ejection sensors (SA1, 7, 13 and 19) are
defective.
The problem is solved, that "NOW
LOADING" is kept appearing to cause
the RU to get unreactive as a plurality of
FTP servers are accessed when the RU
power is turned ON.
The CL name and the IP barcode are
added to the retransmission window on
the operation panel.
The message indicating the error
conditions has been displayed only on the
RU when "Image data signal connector
error" (error code: 11287/11687) occurs.
As measures against this, a modification
is made so that the message prompting
the CL to confirm the RU operation panel
appears.
A modification is made so that the RU
status displayed on the CL is soon
updated when the line between the RU
and the CL recovers from disconnection.
RU software version
Reason
V2.0 V2.1 V2.2 V2.3 V2.4
Changed
Changed
Corrected
Corrected
Changed
Changed
Changed
CR-IR 363 Service Manual
MD-72
Reference
{MD:7.5}
MD-72

Advertisement

Table of Contents
loading

This manual is also suitable for:

Clearview csClearview csmCr-ir 363

Table of Contents