Error
Error Name
Code
Forbidden Response to INQ
12525
command
Forbidden Response to Start-
12526
Read command
Back-side PMT analog power
12611
supply error
Back-side High-voltage command
12617
value error
Back-side High-voltage detection
12618
value error
Back-side scanner control board
12642
error (2)
Back-side photomultiplier control
12651
board error (1)
Back-side photomultiplier control
12652
board error (2)
Image data variation or back-
12687
side image signal connector error
during routine processing
Back-side scanner send/receive
12695
error
Back-side scanner send sum
12696
check error
011-202-11E
06.29.2012 FM6046
Although the INQ command was transmitted from the RU to the CL immediately after cassette insertion during
routine processing, a rejection command was received from the CL. After the processed IP returns to a cassette,
the cassette is ejected so that the associated message appears on the RU panel.
Although the Start-Read command was transmitted from the RU immediately before IP reading during routine
processing, a rejection command was received from the CL. After the processed IP returns to a cassette, the
cassette is ejected so that the associated message appears on the RU panel.
During bootup and routine processing, only error for the PMT analog power supply signal (+15VOKH, +15VOKL) for
back-side is detected.
During bootup and routine processing, error is detected only for the high-voltage command value diagnostics signal
for back-side on the SCT17B board.
During bootup and routine processing, error is detected only for the high-voltage detection value diagnostics signal
for back-side on the SCT17B board.
During routine processing, because the PMR high-voltage value is not set correctly on the SCT17B board, error is
detected for the high-voltage command diagnostics for back side.
During bootup and routine processing, only error for the PMR analog power supply signal (+15VOKH, +15VOKL) is
detected (detail code: 008100). Note that if the HV is OFF, the detail code is 008000.
During bootup and routine processing, only error for the high voltage power supply signal (HVOKH, HVOKL) for
back-side is detected.
An abnormal image data change was detected when the image data transfer diagnostics was conducted for scanner
diagnostics during routine processing.
In the image data transfer diagnostics for scanner diagnostics during routine processing, an abnormal virtual data
change, disconnected image signal connector, or erroneous connector connection was detected.
During bootup, or during routine processing, or in the Maintenance Utility, control communication between the
CPU91A board and SCT17B board is performed; however, because error for the scanner unit is detected, the
control communication does not end normally.
During bootup, during routine processing, or in the Maintenance Utility, control communication between the CPU91A
board and SCT17B board can be done; however, sum check error is detected from the SCT17B board.
Occurrence Condition
CR-IR 363 Service Manual
MT-45
Analysis
Flow
{MT:14.}
{MT:14.}
{MT:5.13}
{MT:5.13}
{MT:5.13}
{MT:5.19}
{MT:5.13}
{MT:5.13}
{MT:5.18}
{MT:5.19}
{MT:5.19}
MT-45