Error
Error Name
Code
Image processing initialization
10100
error
Image processing CPU sequence
10120
error
10230
Scanner functional error
10231
Scanner power supply error (1)
10232
Scanner power supply error (2)
10233
Scanner power supply error (3)
011-202-01
08.25.2003 FM4098
During bootup or during routine processing, when an attempt is made to initialize the image processing section of
the IMG board, 12101 error is detected.
Because it is detected three or more times, fatal error results.
During IP read processing, an attempt is made to activate the sub-CPU (image processing section) of the IMG
board, but 12101 error is detected three times or more, so that fatal error results.
During bootup or during routine processing, because the connector (CN1 on the PMT17A/PMR17B board, CNZ31
on the scanning optics unit, CNZ32 on the scanning optics unit, CN1 on the SYN12A board, CN1 on the LDD17B
board) is disconnected, the laser (LDD), start-point detection (SYN), and photomultiplier (PMT/PMR) functions are
disabled.
Because the laser is not enabled, the leading-edge detection (SED) cannot be performed (detail code: D0E503).
Note that if the HV is OFF, the detail code is D0E403.
During bootup or during routine processing, because +15V is not supplied to the SCN17A/SCT17B board due to
blow of the fuse (G011 on the SCNS17A/CT17B board) or +15V power supply error, most of the functions, except
for the polygonal mirror (POL) and laser (LDD), are disabled (detail code: A00300).
Note that if the HV is OFF, the detail code is A00000.
During bootup or during routine processing, because the fuse (G011 and G041 on the SCN17A/SCT17B board) is
blown, the photomultiplier (PMT/PMR) and start-point detection (SYN) functions are disabled (detail code: 809500).
Note that if the HV is OFF, the detail code is 808400.
During bootup or during routine processing, because the fuse (H011 on the SCN17A/SCT17B board) is blown or
because -15V is not supplied to the SCN17A/SCT17B board, the photomultiplier (PMT/PMR), start-point detection
(SYN), and laser (LDD) functions are disabled.
Because the laser (LDD) is not enabled, the leading-edge detection (SED) cannot be performed (detail code: