FujiFilm PROFECT CS Service Manual page 920

Table of Contents

Advertisement

2
Verify that there is no "unusual nonuniformity" found on the output
film or on the image displayed on the image monitor.
CHECK
u
u
There should be no uneven density difference in the main scan direction.
If there is any uneven density difference, compare the films outputted from the
two IPs of the same size, and determine whether such nonuniformity is caused
by the machine or X-ray tube.
l If there is uneven density difference in the same direction on two films
→ The machine may be the cause of such nonuniformity.
{Troubleshooting}
l If positions of uneven density difference are different
→ The X-ray tube may be the cause of such nonuniformity.
011-202-11E
06.29.2012 FM6046
3
Check the image format.
Make sure that the white blank portion is 2 mm or less, as actually measured on
the IP, from the outermost edge of the image frame of the film outputted.
CHECK
u
u
The white blank portion should be 2 mm or less.
If anything abnormal is found:
{Troubleshooting}
NOTE
u
u
The distance measured on the film may differ from the actual size on the IP,
depending on the reading size and film size. The actual size on the IP should be
calculated using the distance measured on the film and the reduction factor for
the film. Compute the actual size on the IP according to the following equation.
CR-IR 363 Service Manual
MC-431
MC-431