l Setting for warm-up function and dummy cassette detection function
(only for units for use in Japan)
Sets whether or not the warm-up function and the dummy cassette detection function
are used when the machine is started.
- "OFF" (default):
- "WARM-UP & CAS-DETECT": The warm-up function and the dummy cassette
- "CAS-DETECT ONLY":
CAUTION
The warm-up function and the dummy cassette detection function are available
as an option for only the units used in Japan. If the machine is used outside
Japan, always set to "OFF".
l Alarm setting for lamp life
- LOG&MESSAGE:
Written into the log file with an error message displayed.
- LOG ONLY(default): Only written into the log file.
l Alarm setting for erase lamp cleaning
Set the timing to display the alarm for cleaning the erase lamp.
- "NO DISP" (default):
The alarm will not be displayed.
- "90,000":
The alarm is displayed after the erase lamp is lighted up
90,000 times.
- "45,000":
The alarm is displayed after the erase lamp is lighted up
45,000 times.
- "180,000":
The alarm is displayed after the erase lamp is lighted up
180,000 times.
- "FREE INPUT":
Enter a value for "CONDITION OF ERASE LAMP
CLEANING(FREE INPUT)"
l Alarm setting for erase lamp cleaning (free input)
Enter a value for the timing to clean the erasure lamp (only when "FREE INPUT" is
selected at "CONDITION OF ERASE LAMP CLEANING").
011-202-08E
09.30.2009 FM5540
The warm-up function and the dummy cassette
detection function are not available.
detection function are available.
The warm-up function is not available, but the
dummy cassette detection function is available.
l Alarm setting for pump life
Set the timing to display alarm for pump life.
- "NO DISP" (default):
The alarm will not be displayed.
- "90,000":
The alarm is displayed after the pump works 90,000 times.
- "180,000":
The alarm is displayed after the pump works 180,000 times.
- "270,000":
The alarm is displayed after the pump works 270,000 times.
- "FREE INPUT":
Enter a value for "CONDITION OF PUMP LIFE
(FREE INPUT)".
l Alarm setting for pump life (free input)
Enter a value for the timing to display the alarm for lamp life (only when "FREE IN-
PUT" is selected at "CONDITION OF PUMP LIFE").