10234
Scanner power supply error (4)
10235
Scanner power supply error (5)
10236
Scanner power supply error (6)
10244
Scanner control board error (4)
011-202-04
11.15.2004 FM4425
- During bootup or during routine processing, because the fuse (H011 on the SCN17A/SCT17B board) is blown, ±
15V is not supplied to the SCN17A/SCT17B board, and the photomultiplier (PMT/PMR), start-point detection (SYN),
laser (LDD), and SCN board (PLL) functions are disabled. Because the laser (LDD) is not enabled, the leading-
edge detection (SED) cannot be performed (detail code: A00303).
Note that if the HV is OFF, the detail code is A00003.
- During bootup, all the fuses except for the fuse (G04 on the SCN17A/SCT17B board) are blown, ±15V is not
supplied to the SCN17A/SCT17B board, and the photomultiplier (PMT/PMR), start-point detection (SYN), laser
(LDD), and SCN board (PLL) functions are disabled. Because the laser (LDD) is not enabled, the leading-edge
detection (SED) cannot be performed (detail code: B00300).
Note that if the HV is OFF, the detail code is B00000.
- During bootup or during routine processing, because +24V is not supplied to the SCN17A/SCT17B board due to
blow of the fuse (J011 on the SCN17A/SCT17B board) or +24V (SLOT-J) power supply error, the polygonal mirror
(POL) function is disabled (detail code: 902403).
Note that if the HV is OFF, the detail code is 903403.
- During bootup or during routine processing, because +5V is not supplied to the SYN12A board and SED17A board
due to blow of the fuse (A061 or G021 on the SCN17A/SCT17B board) or +5V (SLOT-A) power supply error, the
leading-edge detection (SED) and start-point detection (SYN) functions are disabled (detail code: C01400).
During bootup or during routine processing, because ±15V is not supplied to the SYN12A board, PMT17A/
PMR17B board, and LDD12A board due to blow of the fuses (G011 and H011 on the SCN17A/SCT17B board), the
photomultiplier (PMT/PMR), start-point detection (SYN), and laser (LDD) functions are disabled. Because the laser
(LDD) is not enabled, the leading-edge detection (SED) cannot be performed (detail code: 80D500).
Note that if the HV is OFF, the detail code is 80D400.
During bootup, an SCN17A/SCT17B board diagnostic error associated with polygon error is detected.
Because the PLL on the SCN17A/SCT17B board does not oscillate, a polygon sync signal cannot be detected.
Occurrence Condition
CR-IR 363 Service Manual
MT-13
Analysis
Flow
{MT:5.5}
{MT:5.6}
{MT:5.3}
{MT:5.8}