4.
Enter the main scan length (e.g., "2.8") calculated at step 3.
NOTES
u
u
- After entering the main scan length, do not turn OFF the power of the RU. If
the RU is powered OFF, the main scan length data will not be retained.
- "Input range" only accepts values ranging from -5 to 5.
REFERENCE
The white blank portion of the output image narrows when a negative (-) value is
entered, while it expands when a positive (+) value is entered.
011-202-11E
06.29.2012 FM6046
5.
Make sure that "RESULT-OK" appears on screen.
6.
Using menus "TEST", "Image Format" or "Image Format - 2", and
"SINGLE", read one of the IPs prepared in step 1, and generate an
image output.
u INSTRUCTION u
Be sure to follow the instruction mentioned in "20.2 Image/Conveyance Checks"
when outputting the image.
{MC:20.2_Image/Conveyance Checks}
7.
Measure the actual size of the 175mm-length steel rule and the size of
the white blank portion, and verify that there is no misalignment.
8.
Back up the data so adjusted into a floppy diskette.
{MC:19.1_Backing Up the Scanner Data}