l For "HR-BD" IP
- X-ray tube:
Molybdenum X-ray tube
- Exposure X-ray dose: 20 mR
- Maximum size:
IP of 18"x24" HR-BD or 24"x30" HR-BD size
- Reference conditions: Distance 0.55 m
Voltage 25 kV
Amperage 100 mA
Time 0.053 sec
011-202-11E
06.29.2012 FM6046
20.2
Image/Conveyance Checks
u NOTE u
Because the "MENU" and so forth differ depending on the IP type, image checks
should be performed with the "MENU" specified below.
u INSTRUCTION u
While the "Maintenance is in progress" message is displayed at the center of the LCD
panel, sequentially touch the upper left and upper right corners to allow the LCD panel
to accept touches on the panel.
1.
Using "TEST", "Image Format" or "Image Format-2" and "SINGLE",
record the IP exposed at "20.1 Check Before Procedures", and
generate an image output.
REFERENCE
Refer to "19.2 Checking the Image" for detailed procedures for inputting the
image.
{MC:19.2_Checking the Image}
CR-IR 363 Service Manual
MC-430
MC-430