5.2
Error Code Analysis Flow: Scanner 2
10231
n Analysis Flow
m
Check the error log in the ERROR DB.
Q1
Errors 10230-10236 logged?
Y
Check the error code table again and troubleshoot.
N
{MT:2._Error Code Table}
A
Q2
Check the voltage of the SCN17A/SCT17B board test pin (CN5).
{MT:8._Procedures for Checking the Voltage}
Q3
Check the fuse H011 on the SCN17A/SCT17B board.
{MT:9._Procedures for Checking the Fuses}
Q4
Error recurs?
N
A
Y
Q5
CPU91A and SCN17A/SCT17B boards tested in MUTL normal?
{MT:18._Board Test in MUTL}
N
Q7
Y
Q6
Error recurs?
N
A
Y
m
Disconnect and reconnect the connectors in the following sequence. If the error is
detected, replace the affected part.
1. SCN17A/SCT17B board
{MC:13.4_Scanner Board}
2. CPU91A board
{MC:13.2_CPU91A Board}
011-202-04
11.15.2004 FM4425
Q7
Reseat the error-causing board and check if the error occurs again.
N
A
Y
m
Replace the error-causing board.
{MC:13.2_CPU91A Board}
{MC:13.4_Scanner Board}