When the IP returns to the
cassette, the cassette hold
solenoid is released, and the
cassette hold sensor (SA3, 9,
15 or 21) closes.
The clutch activates to push
out the cassette. The actuator
ascends and the cassette IN
sensor (SA2, 8, 14 or 20) opens.
The cassette removal lamp
(
) lights up.
When the cassette is drawn out,
the movable guide ascends.
The inch/metric sensor (SA6,
12, 18 or 24) opens.
When the cassette is removed,
the shutter closes. The cassette
ejection sensor (SA1, 7, 13 or
19) opens.
011-202-07
09.30.2008 FM5340
REFERENCE
The following displays appear if the status change of the cassette set unit sensors is
checked by means of the monitoring function for operations of setting the cassette on
a shelf on the first stage, reading the IP, and removing the cassette.