FujiFilm PROFECT CS Service Manual page 193

Table of Contents

Advertisement

Error
Error Name
Code
Scanning optics unit board error
12283
during bootup
12284
LD monitor value error warning
12285
HV command value error warning
12286
HV detection value error warning
Image data variation or image
12287
signal connector error during
routine processing
12291
Reading sequence error
Reading sequence error during
12292
routine (2)
12295
Scanner send/receive error
12296
Scanner send sum check error
12297
Communication receive size error
011-202-11E
06.29.2012 FM6046
During bootup, the laser (LDD), start-point detection (SYN), and leading-edge detection (SED) functions are
disabled due to laser failure or disconnected connector.
During IP reading, the laser drive current value (LDIF) becomes more than 1.4 times the factory default value.
During IP reading, error is detected for the HV command value signal sent from the SCN17A/SCT17B board to the
PMT17A/PMR17B board. The difference between MIN and MAX of the monitor value relative to the setup value is
out of specification.
During routine processing, error is detected for the HV detection value signal of the PMT17A/PMR17B board. The
difference between MIN and MAX of the monitor value relative to the setup value is out of specification.
An abnormal image data change was detected when the image data transfer diagnostics was conducted for scanner
diagnostics during routine processing.
In the image data transfer diagnostics for scanner diagnostics during routine processing, an abnormal virtual data
change, disconnected image signal connector, or erroneous connector connection was detected.
During routine processing, error for the VSYNC signal on the SCN17A/SCT17B board is detected after IP reading is
completed (detail code: 000002).
Reading is completed by any interrupt other than the leading-edge detection interrupt or end-of-screen interrupt.
At the same time, an insufficient number of output lines and a too-early start of reading are detected (detail code:
000003).
During bootup, or during routine processing, or in the Maintenance Utility, control communication between the
CPU91A board and SCN17A/SCT17B board is performed; however, because error for the scanner unit is detected,
the control communication does not end normally.
During bootup, during routine processing, or in the Maintenance Utility, control communication between the CPU91A
board and SCN17A/SCT17B board can be done; however, sum check error is detected from the SCN17A/SCT17B
board.
During bootup, or during routine processing, or in the Maintenance Utility, communication between the CPU91A
board and SCN17A/SCT17B board can be done; however, data size error is detected from the SCN17A/SCT17B
board.
Occurrence Condition
CR-IR 363 Service Manual
MT-36
Analysis
Flow
{MT:5.15}
{MT:5.15}
{MT:5.16}
{MT:5.16}
{MT:5.17}
{MT:18.}
-
{MT:5.19}
{MT:5.19}
{MT:5.19}