Hide thumbs Also See for CR-IR346RU:

Advertisement

Quick Links

CR-IR346RU Service Manual
Installation (IN)
(Philips Medical Systems 01.1)

Advertisement

Table of Contents
loading

Summary of Contents for FujiFilm CR-IR346RU

  • Page 1 CR-IR346RU Service Manual Installation (IN) (Philips Medical Systems 01.1)
  • Page 2: Table Of Contents

    CR-IR346RU Service Manual – Contents Installation (IN) Specifications of Machine ................... IN-2 Installation Work Flowchart ..................IN-4 Preparation for Installation ..................IN-6 3.1 Precautions Regarding Installation..............IN-6 3.2 Unloading ......................IN-7 3.3 Transfer....................... IN-11 3.4 Temporary Placement ..................IN-12 3.5 Checking the Items Supplied ................IN-13 Installation Procedures ....................
  • Page 3 Control Sheet Issue date Revision number Reason Pages affected 10/10/2000 New release (FM2887) All pages 08/30/2001 Support for software version A05 (FM3058) IN-5, 6, 12–14, 17–19, 21.1–21.2, 23, 24–29, 31–35, 38 010-051-01 CR-IR346RU Service Manual IN - 1 08.30.2001 FM3058...
  • Page 4: Specifications Of Machine

    • Moving : Double-wheel caster (variable-direction/no-brake) x4 • Fixing : • Adjustable foot x4 • Machine retainer x2 • Table x1 (optional) • Table x1 and table retainer x2 (optional) 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 2 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
  • Page 5 G Minimum space for table (full option) installation Machine Machine Table Table 1,000 1,000 ✻ ✻ Make sure that servicing space is secured. Unit: mm Make sure that servicing space is secured. FR6H7004.EPS 010-051-00 CR-IR346RU Service Manual IN - 3 10.10.2000 FM2887 (1)
  • Page 6: Installation Work Flowchart

    5.1 Assembling the Table 6. Setting the CPU12A Board 7. Connecting the Cables 7.1 Connecting the Power Cable and Checking Resistance Value 7.2 Connecting the Interface Cable FR6H7001.EPS 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 4 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
  • Page 7 11. Powering ON the CL/RU 12. Image/Conveyance Checks 12.1 Conveyance Checks 12.2 Image Checks 13. Powering OFF the CL/RU 14. Cleaning the CL/RU 15. Attaching the Cassette Handling Precaution Label FR6H7005.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 5 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 8: Preparation For Installation

    When servicing a board or optical unit, use a wristband to ground your body. If you body is not grounded, the electrostatic buildup on your body may damage the electronic parts. 010-051-00 CR-IR346RU Service Manual IN - 6 010-051-01 10.10.2000 FM2887 (1)
  • Page 9: Unloading

    [Remove] Top machine-wrapping protective material PP band [Remove] (x4) [Remove] Tray DO NOT ROLL UNSTABLE VORSICHT [Remove] Slope #5 [Remove] Machine-wrapping protective material Machine-wrapping protective material Carrier pallet FR6H7E01.EPS 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 7 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
  • Page 10 If the machine is unpacked immediately after it is carried into the room, moisture condensation may occur on the machine. (3) Remove the vinyl cover. Machine [Remove] Vinyl cover Carrier pallet FR6H7652.EPS 010-051-00 CR-IR346RU Service Manual IN - 8 10.10.2000 FM2887 (1)
  • Page 11 [Remove] Displacement prevention member FR6H7602.EPS (5) Attach the displacement prevention member. Fitting [Secure] Na6 Displacement [Remove] Na6 prevention member Slope Slope [Shift] Fitting Na6 removed at #1 should be used at #3. FR6H7619.EPS 010-051-00 CR-IR346RU Service Manual IN - 9 10.10.2000 FM2887 (1)
  • Page 12 IN - 10 (6) Slide the position of the fitting of the slope. [Attach] Slope FR6H7606.EPS REFERENCE The slope may be attached only to the front side of the machine. 010-051-00 CR-IR346RU Service Manual IN - 10 10.10.2000 FM2887 (1)
  • Page 13: Transfer

    Note that the step over which the machine may move is about 10 mm high at most. (1) Transfer the machine into the installation place. 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 11 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
  • Page 14: Temporary Placement

    1,000 mm or greater in space is secured on one side of the machine. (2) Secure the machine with the adjustable feet in place. Machine [Secure] Na12 (x4) Caster Adjustable foot FR6H7013.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 12 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 15: Checking The Items Supplied

    One set (Inch/metric label) Fuse set Spare Ferrite bead E04SR301334 Used only for direct connection CL/RU connection cable without hub Category 5 cable (cross, UTP type), 1.95 m TR6H7010.EPS 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 13 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 16 Retainer For retaining the machine • 100-120 V: U.S., Canada, etc. • 200-230 V: Germany, France, RU power cable Spain, Sweden, etc. • 240 V: U.K. TR6H7011.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 14 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 17: Installation Procedures

    IN - 15 Installation Procedures Removing the Machine-Wrapping Packing Materials (1) Remove the packing materials. [Remove] Tape [Remove] Retaining member [Remove] Tape [Remove] Tape [Remove] Tape (x2)/front cover [Remove] Tape FR6H7607.EPS 010-051-00 CR-IR346RU Service Manual IN - 15 10.10.2000 FM2887 (1)
  • Page 18: Removing The Light Protect Plates

    IN - 16 Removing the Light Protect Plates (1) Remove the light protect plates. [Remove] T4x6 (x6) Upper light protect plate FR6H7657.EPS 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 16 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
  • Page 19: Repositioning The Retaining Members

    One screw for each of the right and left retaining members removed at step #2 should be used at step #5. (2) Attach the upper light protect plate. “4.2 Removing the Light Protect Plates” 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 17 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 20: Installing The Options

    1. Pillar (left) 2. Pillar (right) 3. Rear pipe 4. Top plate 5. Shelf 6. Bracket 7. T6x35 (x8) 8. T6x12 (x4) 9. T4x6 (x4) FR6H7701.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual (PMS _01.1) IN - 18 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 21 G Parts used to secure the machine 12. Bracket 13. Retainer (x2) 14. Na12 (x4) 15. T6x12 (x2) 16. Bolt 12x20 (x4), W12 (x4) 17. Bolt 10x20 (x4) FR6H7710.EPS 010-051-00 010-051-01 CR-IR346RU Service Manual (PMS 01.1) IN - 19 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 22 I Assembly Procedures (1) Attach the rear pipe. Rear pipe [Secure] T6x35 (x8) Pillar (left) Pillar (right) FR6H7717.EPS (2) Attach the top plate. Top plate [Secure] T6x12 (x4) FR6H7718.EPS 010-051-00 CR-IR346RU Service Manual (PMS _01.1) IN - 20 10.10.2000 FM2887 (1)
  • Page 23 IN - 21 (3) Attach the bracket. [Secure] T4x6 (x4) Bracket FR6H7719.EPS (4) Attach the shelf. [Attach] Shelf Shelf FR6H7720.EPS 010-051-00 CR-IR346RU Service Manual (PMS_0.1) IN - 21 10.10.2000 FM2887 (1)
  • Page 24 IN - 21.1 (6) Attach the shock absorbing materials. Pillar (right) Pillar (left) Shock Shock absorbing absorbing material material 10mm 10mm FR6H7749.EPS (6) Attach the covers. [Insert] Cover (x10) FR6H7750.EPS 010-051-01 CR-IR346RU Service Manual (PMS_01.1) IN - 21.1 08.30.2001 FM3058...
  • Page 25 IN - 21.2 BLANK PAGE 010-051-01 CR-IR346RU Service Manual IN - 21.2 08.30.2001 FM3058...
  • Page 26: Setting The Cpu12A Board

    Before handling the board, wear the wristband. Inch cassettes require a HW change. (1) Remove the lower light protect plate. Lower light protect plate [Remove] T4x6 (x4) FR6H7658.EPS 010-051-00 010-051-00 CR-IR346RU Service Manual (PMS_01.1) IN - 22 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
  • Page 27 For design analysis – – OFF (fixed) S2 settings Description Remote Local Default Interlink with CL – – Local (fixed) FR6H7659.EPS (3) Attach the lower light protect plate. 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 23 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 28: Connecting The Cables

    <For use in Japan and the US> <For use in Europe> <For use in the UK, etc.> FR6H7007.EPS (4) Place the breaker switch in the OFF position. 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 24 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 29: Connecting The Interface Cable

    • Make sure that 50 to 80 mm of the cable protrudes from the ferrite bead (on the side you will install into the RU). 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 25 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 30: Direct Connection Of The Cl/Ru

    (2) Connect the I/F cable to the I/F cable connector of the RU. REFERENCE The I/F cable should be connected with its ferrite bead end on the RU side. Ferrite bead I/F cable connector I/F cable FR6H7725.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 26 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 31: Network Connection

    The I/F cable should be connected with its ferrite bead end on the RU side. Ferrite bead I/F cable connector I/F cable Power cable Ferrite bead I/F cable Switching hub (100 BASE - T or 10 BASE - T) Network FR6H7725.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 27 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 32: Installing The Cover

    IN - 28 Installing the Cover (1) Install the front cover. [Install] Front cover FR6H7801.EPS CL Installation Procedures (1) Perform the installation procedures of the other system components. 010-051-00 010-051-01 CR-IR346RU Service Manual (PMS_01.1) IN - 28 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 33: Final Placement

    • When the adjustable feet are secured, the casters should be detached from the floor. (2) Secure the adjustable feet. Machine [Secure] Na12 (x4) Caster Adjustable foot FR6H7724.EPS 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 29 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 34: Securing The Machine With The Retainers

    Retainer Spacer Adjustable foot FR6H7600.EPS (1) Determine the locations for embedding the anchor nuts. 614 mm (reference value) Machine 398 mm (reference value) Marking location Retainer FR6H7613.EPS 010-051-00 CR-IR346RU Service Manual IN - 30 10.10.2000 FM2887 (1)
  • Page 35 • When the adjustable feet are secured, the casters should be detached from the floor. [Secure] Bolt 12x20, W12 (x2) [Move] Machine [Secure] Bolt 12x20, W12 (x2) [Secure] Adjustble foot Na12 (x4 or x2) FR6H7614.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 31 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 36: Securing The Table And Machine Together

    (1) Secure the machine to the table. Table [Move] Machine [Secure] T6x12 (x2) Bracket [Secure] Adjustable foot Bracket Bracket [Secure] Na12 (x4 or x2) Adjustable foot (machine) FR6H7617.EPS 010-051-00 010-051-01 CR-IR346RU Service Manual (PMS_01.1) IN - 32 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
  • Page 37: Securing The Machine With The Retainers And Table

    Before drilling holes for embedding the anchor nuts, wear protective goggles. REFERENCE When determining the locations for embedding the anchor nuts, perform temporary installation with the retainers attached to the table, and then mark the locations. 010-051-01 010-051-00 CR-IR346RU Service Manual (PMS_01.1) IN - 33 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 38 18φ #1 [Attach]/#3 [Remove] Bolt 12x20 (x4) W12 (x4) Anchor nut Na12 52mm FR6H7723.EPS (4) Secure the table in place. Table [Secure] Bolt 12x20 (x4) FR6H7615.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual (PMS_01.1) IN - 34 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 39 (5) Secure the machine to the table. Table [Move] Machine [Secure] T6x12 (x2) Bracket Bracket Bracket [Secure] Na12 (x4 or x2) Adjustable foot (machine) FR6H7616.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual (PMS_01.1) IN - 35 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 40: Powering On The Cl/Ru

    IN - 36 11. Powering ON the CL/RU (1) Power ON either the Console or the Preview Unit, User Terminal and Easy Vision. FR6H7747.EPS (2) Power ON the RU. FR6H7746.EPS 010-051-00 CR-IR346RU Service Manual (PMS 01.1) IN - 36 10.10.2000 FM2887 (1)
  • Page 41: Image/Conveyance Checks

    Note, however, that those conditions are for standard IP setup reference. Distance: 1.8 m Voltage: 80 kV Amperage: 50 mA Time: 0.013 sec 010-051-00 CR-IR346RU Service Manual IN - 37 10.10.2000 FM2887 (1)
  • Page 42: Conveyance Checks

    (1) Verify that there is no unusual nonuniformity found on the output film and on the image displayed on the image monitor. N CHECK N There should be no unusual nonuniformity. If anything abnormal is found, see [Troubleshooting Volume]. 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 38 08.30.2001 FM3058 10.10.2000 FM2887 (1)
  • Page 43: Powering Off The Cl/Ru

    (2) Power OFF the RU. FR6H7656.EPS 14. Cleaning the CL/RU (1) Clean the monitor and covers of the CL with a dry cloth. (2) Clean the covers of the RU with a moistened cloth. 010-051-00 CR-IR346RU Service Manual IN - 39 10.10.2000 FM2887 (1)
  • Page 44: Attaching The Cassette Handling Precaution Label

    (1) Attach the cassette handling precaution label above the cassette size indication. MAKE SURE TO REMOVE ALL EXPOSURE MARKERS USED AT THE TIME OF EXPOSURE BEFORE INSERTING THE CASSETTE. [Attach] Label FR6H7644.EPS 010-051-00 CR-IR346RU Service Manual IN - 40 10.10.2000 FM2887 (1)
  • Page 45 IN - 41 BLANK PAGE 010-051-00 CR-IR346RU Service Manual IN - 41 10.10.2000 FM2887 (1)

Table of Contents