CR-IR346RU Service Manual – Contents Installation (IN) Specifications of Machine ................... IN-2 Installation Work Flowchart ..................IN-4 Preparation for Installation ..................IN-6 3.1 Precautions Regarding Installation..............IN-6 3.2 Unloading ......................IN-7 3.3 Transfer....................... IN-11 3.4 Temporary Placement ..................IN-12 3.5 Checking the Items Supplied ................IN-13 Installation Procedures ....................
Page 3
Control Sheet Issue date Revision number Reason Pages affected 10/10/2000 New release (FM2887) All pages 08/30/2001 Support for software version A05 (FM3058) IN-5, 6, 12–14, 17–19, 21.1–21.2, 23, 24–29, 31–35, 38 010-051-01 CR-IR346RU Service Manual IN - 1 08.30.2001 FM3058...
Page 5
G Minimum space for table (full option) installation Machine Machine Table Table 1,000 1,000 ✻ ✻ Make sure that servicing space is secured. Unit: mm Make sure that servicing space is secured. FR6H7004.EPS 010-051-00 CR-IR346RU Service Manual IN - 3 10.10.2000 FM2887 (1)
5.1 Assembling the Table 6. Setting the CPU12A Board 7. Connecting the Cables 7.1 Connecting the Power Cable and Checking Resistance Value 7.2 Connecting the Interface Cable FR6H7001.EPS 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 4 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
Page 7
11. Powering ON the CL/RU 12. Image/Conveyance Checks 12.1 Conveyance Checks 12.2 Image Checks 13. Powering OFF the CL/RU 14. Cleaning the CL/RU 15. Attaching the Cassette Handling Precaution Label FR6H7005.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 5 08.30.2001 FM3058 10.10.2000 FM2887 (1)
When servicing a board or optical unit, use a wristband to ground your body. If you body is not grounded, the electrostatic buildup on your body may damage the electronic parts. 010-051-00 CR-IR346RU Service Manual IN - 6 010-051-01 10.10.2000 FM2887 (1)
[Remove] Top machine-wrapping protective material PP band [Remove] (x4) [Remove] Tray DO NOT ROLL UNSTABLE VORSICHT [Remove] Slope #5 [Remove] Machine-wrapping protective material Machine-wrapping protective material Carrier pallet FR6H7E01.EPS 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 7 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
Page 10
If the machine is unpacked immediately after it is carried into the room, moisture condensation may occur on the machine. (3) Remove the vinyl cover. Machine [Remove] Vinyl cover Carrier pallet FR6H7652.EPS 010-051-00 CR-IR346RU Service Manual IN - 8 10.10.2000 FM2887 (1)
Page 11
[Remove] Displacement prevention member FR6H7602.EPS (5) Attach the displacement prevention member. Fitting [Secure] Na6 Displacement [Remove] Na6 prevention member Slope Slope [Shift] Fitting Na6 removed at #1 should be used at #3. FR6H7619.EPS 010-051-00 CR-IR346RU Service Manual IN - 9 10.10.2000 FM2887 (1)
Page 12
IN - 10 (6) Slide the position of the fitting of the slope. [Attach] Slope FR6H7606.EPS REFERENCE The slope may be attached only to the front side of the machine. 010-051-00 CR-IR346RU Service Manual IN - 10 10.10.2000 FM2887 (1)
Note that the step over which the machine may move is about 10 mm high at most. (1) Transfer the machine into the installation place. 010-051-00 010-051-00 CR-IR346RU Service Manual IN - 11 10.10.2000 FM2887 (2) 10.10.2000 FM2887 (1)
1,000 mm or greater in space is secured on one side of the machine. (2) Secure the machine with the adjustable feet in place. Machine [Secure] Na12 (x4) Caster Adjustable foot FR6H7013.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 12 08.30.2001 FM3058 10.10.2000 FM2887 (1)
One set (Inch/metric label) Fuse set Spare Ferrite bead E04SR301334 Used only for direct connection CL/RU connection cable without hub Category 5 cable (cross, UTP type), 1.95 m TR6H7010.EPS 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 13 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
Page 16
Retainer For retaining the machine • 100-120 V: U.S., Canada, etc. • 200-230 V: Germany, France, RU power cable Spain, Sweden, etc. • 240 V: U.K. TR6H7011.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 14 08.30.2001 FM3058 10.10.2000 FM2887 (1)
One screw for each of the right and left retaining members removed at step #2 should be used at step #5. (2) Attach the upper light protect plate. “4.2 Removing the Light Protect Plates” 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 17 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
<For use in Japan and the US> <For use in Europe> <For use in the UK, etc.> FR6H7007.EPS (4) Place the breaker switch in the OFF position. 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 24 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
• Make sure that 50 to 80 mm of the cable protrudes from the ferrite bead (on the side you will install into the RU). 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 25 08.30.2001 FM3058 10.10.2000 FM2887 (1)
(2) Connect the I/F cable to the I/F cable connector of the RU. REFERENCE The I/F cable should be connected with its ferrite bead end on the RU side. Ferrite bead I/F cable connector I/F cable FR6H7725.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 26 08.30.2001 FM3058 10.10.2000 FM2887 (1)
The I/F cable should be connected with its ferrite bead end on the RU side. Ferrite bead I/F cable connector I/F cable Power cable Ferrite bead I/F cable Switching hub (100 BASE - T or 10 BASE - T) Network FR6H7725.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 27 08.30.2001 FM3058 10.10.2000 FM2887 (1)
IN - 28 Installing the Cover (1) Install the front cover. [Install] Front cover FR6H7801.EPS CL Installation Procedures (1) Perform the installation procedures of the other system components. 010-051-00 010-051-01 CR-IR346RU Service Manual (PMS_01.1) IN - 28 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
• When the adjustable feet are secured, the casters should be detached from the floor. (2) Secure the adjustable feet. Machine [Secure] Na12 (x4) Caster Adjustable foot FR6H7724.EPS 010-051-00 010-051-01 CR-IR346RU Service Manual IN - 29 10.10.2000 FM2887 (1) 08.30.2001 FM3058...
Retainer Spacer Adjustable foot FR6H7600.EPS (1) Determine the locations for embedding the anchor nuts. 614 mm (reference value) Machine 398 mm (reference value) Marking location Retainer FR6H7613.EPS 010-051-00 CR-IR346RU Service Manual IN - 30 10.10.2000 FM2887 (1)
Page 35
• When the adjustable feet are secured, the casters should be detached from the floor. [Secure] Bolt 12x20, W12 (x2) [Move] Machine [Secure] Bolt 12x20, W12 (x2) [Secure] Adjustble foot Na12 (x4 or x2) FR6H7614.EPS 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 31 08.30.2001 FM3058 10.10.2000 FM2887 (1)
Before drilling holes for embedding the anchor nuts, wear protective goggles. REFERENCE When determining the locations for embedding the anchor nuts, perform temporary installation with the retainers attached to the table, and then mark the locations. 010-051-01 010-051-00 CR-IR346RU Service Manual (PMS_01.1) IN - 33 08.30.2001 FM3058 10.10.2000 FM2887 (1)
IN - 36 11. Powering ON the CL/RU (1) Power ON either the Console or the Preview Unit, User Terminal and Easy Vision. FR6H7747.EPS (2) Power ON the RU. FR6H7746.EPS 010-051-00 CR-IR346RU Service Manual (PMS 01.1) IN - 36 10.10.2000 FM2887 (1)
Note, however, that those conditions are for standard IP setup reference. Distance: 1.8 m Voltage: 80 kV Amperage: 50 mA Time: 0.013 sec 010-051-00 CR-IR346RU Service Manual IN - 37 10.10.2000 FM2887 (1)
(1) Verify that there is no unusual nonuniformity found on the output film and on the image displayed on the image monitor. N CHECK N There should be no unusual nonuniformity. If anything abnormal is found, see [Troubleshooting Volume]. 010-051-01 010-051-00 CR-IR346RU Service Manual IN - 38 08.30.2001 FM3058 10.10.2000 FM2887 (1)
(2) Power OFF the RU. FR6H7656.EPS 14. Cleaning the CL/RU (1) Clean the monitor and covers of the CL with a dry cloth. (2) Clean the covers of the RU with a moistened cloth. 010-051-00 CR-IR346RU Service Manual IN - 39 10.10.2000 FM2887 (1)
(1) Attach the cassette handling precaution label above the cassette size indication. MAKE SURE TO REMOVE ALL EXPOSURE MARKERS USED AT THE TIME OF EXPOSURE BEFORE INSERTING THE CASSETTE. [Attach] Label FR6H7644.EPS 010-051-00 CR-IR346RU Service Manual IN - 40 10.10.2000 FM2887 (1)
Page 45
IN - 41 BLANK PAGE 010-051-00 CR-IR346RU Service Manual IN - 41 10.10.2000 FM2887 (1)