FujiFilm PROFECT CS Service Manual page 208

Table of Contents

Advertisement

Error
Error Name
Code
12908
[FRONT]S-value fluctuations (1)
12909
[FRONT]S-value fluctuations (2)
12910
[BACK]S-value fluctuations (1)
12911
[BACK]S-value fluctuations (2)
Touch panel calibration file load
13032
log
13034
UI debug information
initialization from the warm-up
13035
screen in mobile type equipment
Timer value changed from the
13036
warm-up screen in mobile type
equipment
13200
High-voltage OFF during bootup
13210
Polygon timeout
13227
Reading IP size error
011-202-11E
06.29.2012 FM6046
The error was logged because S-value was fluctuated over 50% from initial value.
The error was logged because S-value was fluctuated over 70% from initial value.
The error was logged because S-value was fluctuated over 50% from initial value.
The error was logged because S-value was fluctuated over 70% from initial value.
<Service>
- Because the touch panel is not calibrated after replacement of the IMG board and LCD panel, the file cannot be
loaded.
- During bootup, the touch panel calibration file cannot be loaded from the FLASH ROM to the SD RAM.
- File loading could not be accomplished because no touch panel correction was provided after the RU application
restoration procedure was performed.
Unlikely error condition associated with the RU software occurs.
While the use of mobile-type equipment was selected for configuration setup, the user pressed the "immediate start"
key from the "warm-up" screen, which appeared at startup, to complete startup with the warm-up time reduced.
While the use of mobile-type equipment was selected for configuration setup, the user performed a timer value
change procedure from the "warm-up" screen, which appeared at startup.
During bootup, the photomultiplier (PMT17A/PMR17B board) high voltage is OFF.
During bootup and routine processing, the polygon OK signal is not detected within 7 seconds after turn-ON of the
polygonal mirror.
The IP of the size other than the input size is inserted for shading correction of the M-Utility, so that error is detected.
Occurrence Condition
CR-IR 363 Service Manual
MT-51
Analysis