Error
Error Name
Code
12331
IP-feed air leak warning
12332
Feed IP conveyance error (1)
12333
Feed IP conveyance error (2)
12334
Feed IP conveyance error (3)
12335
Feed IP conveyance error (4)
Feed IP sensor failure (1)
12336
(SB5 failure)
Feed IP sensor failure (2)
12337
(SG1 failure)
12343
IP barcode reading retry
12347
Reading and conveyance retry
011-202-11E
06.29.2012 FM6046
During routine processing/M-Utility/abnormality processing, the IP leak valve (SVA1/2/3/4) is driven to convey the
IP from the cassette to the feed/load conveyor unit, and the IP suction pump (PA1/2/3/4) is stopped; however, the
suction sensor (SA5/11/17/23) does not transition to CLOSE within the specified period of time.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from the
feed/load conveyor unit to the side-positioning conveyor unit; however, the pre-BCR IP sensor (SB5) does not
transition to CLOSE within the specified period of time. Therefore, a retry was performed.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from the
feed/load conveyor unit to the side-positioning conveyor unit; however, the side-positioning entrance IP sensor (SG1)
does not transition to CLOSE within the specified period of time.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from the
feed/load conveyor unit to the side-positioning conveyor unit; however, the pre-BCR IP sensor (SB5) does not
transition to OPEN within the specified period of time. Therefore, a retry was performed.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from the
feed/load conveyor unit to the side-positioning conveyor unit; however, the side-positioning IP sensor (SC3) does
not transition to CLOSE within the specified period of time. Therefore, a retry was performed.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from the
feed/load conveyor unit to the side-positioning conveyor unit; however, the side-positioning IP sensor (SC3) or
side-positioning entrance IP sensor (SG1) transitions to CLOSE before the pre-BCR IP sensor (SB5) transitions to
CLOSE. Therefore, a retry was performed.
During routine processing/M-Utility, the IP transport motor (MB1, MB2, MC3) is driven to convey the IP from
the feed/load conveyor unit to the side-positioning conveyor unit; however, the side-positioning IP sensor (SC3)
transitions to CLOSE before the side-positioning entrance IP sensor (SG1) transitions to CLOSE. Therefore, a retry
was performed.
During routine processing, the barcode cannot be read by the IP barcode reader (BCRC1) of the side-positioning
conveyor unit.
As the IP cannot be conveyed to the reader unit within a specified time, a retry is made.
Occurrence Condition
CR-IR 363 Service Manual
MT-38
Analysis
Flow
{MT:4.41}
{MT:4.23}
{MT:4.23}
{MT:4.23}
{MT:4.23}
{MT:4.23}
{MT:4.23}
{MT:4.28}
-
MT-38