10.20.1 Check Procedures via Monitoring Function (16)
1.
Start the RU PC-TOOL.
{MU:4.2.1_Starting RU PC-TOOL}
2.
Start monitoring of the RU PC-TOOL.
{MT:10.2.1_Starting the Monitoring Function}
3.
Allow the hold release arm to move, and check the monitoring display.
l GOOD indication
011-202-04
11.15.2004 FM4425
l NO GOOD indication
Neither "OPEN" nor "CLOSE" is displayed for the sensor to be checked, after the
hold release arm is moved.
u NOTE u
Click on the [START] button in the monitoring window to initiate the monitoring
beforehand. Unless the monitoring is initiated, nothing is displayed in the window
even when the sensor changes its status.
4.
Exit the monitoring of the RU PC-TOOL.
{MT:10.2.2_Exiting the Monitoring Function}