FujiFilm PROFECT CS Service Manual page 188

Table of Contents

Advertisement

Error
Error Name
Code
11791
Erasure lamp cleaning end 1
11905
IP suction pump life end
11908
[FRONT]S-value fluctuations (1)
11909
[FRONT]S-value fluctuations (2)
11910
[BACK]S-value fluctuations (1)
11911
[BACK]S-value fluctuations (2)
12030
Bitmap file load error
12071
FTP server access error
12072
History file restore
MUTL mode multiple access
12080
warning
12101
Insufficient image data
Transfer retry due to insufficient
12102
image data
12103
Undefined IMG result status
12148
Image frame clock timeout
011-202-11E
06.29.2012 FM6046
The life end warning message appeared because a predetermined cumulative erasure lamp illumination time was
exceeded.
The message appeared and the error was logged because the life end of the IP suction pump was reached.
Replace the pump.
The message appeared and the error was logged because S-value was fluctuated over 50% from initial value.
The message appeared and the error was logged because S-value was fluctuated over 70% from initial value.
The message appeared and the error was logged because S-value was fluctuated over 50% from initial value.
The message appeared and the error was logged because S-value was fluctuated over 70% from initial value.
During bootup, loading of the bitmap file from the FLASH ROM to SD RAM fails.
During bootup and during routine processing, the FTP server cannot be accessed for communication from the RU to
the CL. Or, the FTP server may be accessed, but a connection cannot be established.
As the history file on the FTP server is destroyed, it is restored from the backup file.
- "MUTL mode" functions are simultaneously activated for the RU.
- The function that is entered first is activated, but the function that is entered second is not activated, and error is
logged.
Image data from the IMG17A/IMG17B board is insufficient. Although transfer was retried, a retry-over occurs. The
insufficient data is complemented with image data of a uniform density.
As the image data from the IMG17A/IMG17B board is insufficient, transfer was retried.
Undefined result status was returned from the IMG17A/IMG17B board.
During routine processing, a timeout of the frame clock lock (information indicative of the leading edge of the image
data) sent from the SCT17B board is detected by the sub-CPU. (30 seconds)
Occurrence Condition
CR-IR 363 Service Manual
MT-31
Analysis
Flow
-
-
-
-
-
-
{MT:18.}