21.
Check "RU VERSION", and click on [OK].
22.
Confirm the installation precautions.
CAUTIONS
- While data is being written into the FLASH ROM, never turn OFF the power
of the RU. If the power of the RU is turned OFF, the program residing in the
memory is corrupted, so that the RU cannot be rebooted
- The post-installation checkout procedure varies with the version of the RU
software. Before installation, confirm the version of the RU software that is
installed in the flash ROM on the IMG board.
23.
Click on [OK].
l When the version of the RU software installed in the flash ROM on the IMG
board is 1.3 or earlier
011-202-04
11.15.2004 FM4425
l When the version of the RU software installed in the flash ROM on the IMG
board is 1.4 or later
REFERENCE
The installation process screen appears on the CL display.
CR-IR 363 Service Manual
Appx MT-6
Appx MT-6