Error
Error Name
Code
11366
Load IP suction error (1)
11367
Load IP suction error (2) (drop)
11425
Initialization cassette status error
Initialization IP position
11459
information error
11490
IP is overexposed
11493
Dummy cassette ejection request
11494
Dummy cassette request
11495
Remove an empty cassette
11496
Unread IP was ejected
Empty the largest cassette and
11498
insert the cassette
11499
IP reading suspended
11510
Unread IP ejection (1)
011-202-11E
06.29.2012 FM6046
During bootup/routine processing/M-Utility/abnormality processing, the IP suction pump (PA1/2/3/4) is driven to load
the IP to the cassette; however, the suction sensor (SA5/11/17/23) does not transition to OPEN within the specified
period of time. Though the retries (twice) are executed, the IP cannot be sucked.
During bootup/routine processing/M-Utility, the IP is sucked to perform suction conveyance of the IP from the
feed/load conveyor unit to the cassette; however, the CLOSE state (IP drop) of the suction sensor (SA5/11/17/23)
is detected during IP conveyance. Though the retries (twice) for the IP suction are executed, the IP drop is detected
again.
During bootup, the IP is detected in the cassette set unit; however, the OPEN state of the cassette IN sensor
(SA2/8/14/20) or the CLOSE state of the cassette hold sensor (SA3/9/15/21) is detected for the confirmation of the
presence of cassette, so that inconsistency of the sensor status is detected.
During bootup, the CMOS information (IP found) of the machine does not coincide with the remaining IP search
processing result (IP not found), so that the user is notified.
An overexposed IP was detected during bootup or routine processing.
During bootup, a dummy cassette is detected, and ejection of the cassette is requested.
As no dummy cassette is detected upon termination, and insertion of the dummy cassette is requested.
An empty cassette was detected during bootup.
A cassette was loaded while the CL was not ready for reading.
<User operation>
The IP size was unknown or the ejection destination cassette was not found during bootup.
<Action>
Insert an empty cassette.
A remaining IP that could not be returned as an unprocessed IP was detected at a position behind the subscanning
unit during bootup.
The error code is displayed in the LCD but not logged in the ERROR DB.
<Faulty phenomenon>
The IP can be conveyed during secondary erasure but is ejected to the cassette without being read during routine
processing.
<Occurrence conditions>
This error occurs if CL CONFIG setup is performed to select "0: 1-1" for "21. Connecting Between Reader and
Console" under "1. IMAGE MODALITY".
<Servicing>
The IP is ejected into a cassette without being read due to a cause attributable to the CL during routine processing.
The IP is conveyed to the side-positioning conveyor and returned to the cassette without being read.
Occurrence Condition
CR-IR 363 Service Manual
MT-28
Analysis
Flow
{MT:4.26}
{MT:4.26}
{MT:4.21}
{MT:4.36}
-
-
-
-
-
-
{MT:4.37}
-