FujiFilm PROFECT CS Service Manual page 186

Table of Contents

Advertisement

Error
Error Name
Code
11511
Patient information not registered
11512
ID information obtaining error
Invalid response from Start-Read
11527
command CL
11530
Image data retransmission
11531
Retransmission retry failure
Image data variation or back-
11687
side image signal connector error
during bootup
Insufficient image erasure
11700
because some of the erasure
lamps were unlit.
Erasure time extension due to
11701
erasure lamp illumination failure
Erasure lamp life end advance
11702
notice
11703
Erasure lamp life end 1
11704
Erasure lamp life end 2
011-202-11E
06.29.2012 FM6046
<User operation>
This error occurs during image output when the menu is not selected .
<Occurrence condition>
The patient information corresponding to the barcode of the processed IP has not been registered in the CL.
Alternatively, the barcode cannot be read because the barcode affixed on the IP is soiled or the like.
Although the RU sent an ID information inquiry to the CL during routine processing, the RU could not obtain ID
information due to a CL abnormality (except when the ID information could not be acquired due to an unregistered
ID or barcode error). After the processed IP returns to a cassette, the cassette is ejected so that the associated
message appears on the panel.
Although the CL receives the Start-Read command from the RU, the CL transmits the reject command to the RU
and displays the message accordingly on the CL window.
The RU has retransmitted the image data to the CL
During routine processing, an attempt is made to retransmit the unsent image data from the RU to the CL after
communication with the CL is suspended; however, the CL rejects it, so that the image data is lost and the cassette
is ejected.
An abnormal virtual data change was detected when the virtual data transfer diagnostics was conducted for bootup.
In the virtual data transfer diagnostics for bootup, an abnormal virtual data change, disconnected image signal
connector, or erroneous connector connection was detected.
Insufficiently erased IP was ejected to the cassette because some of the image erasure lamps were unlit.
To confirm the number of unlit lamps, see Chapter MD.
The system turned to erasure degeneration mode because the specified number of image erasure lamps went unlit,
which resulted in ejecting the insufficiently erased IP to the cassette.
A predetermined cumulative erasure lamp illumination time was exceeded. Therefore, the message appeared to
indicate that the lamp life end is about to be reached.
The life end warning message appeared because a predetermined cumulative erasure lamp illumination time was
exceeded.
The life end message appeared because a predetermined cumulative erasure lamp illumination time was exceeded.
Occurrence Condition
CR-IR 363 Service Manual
MT-29
Analysis
Flow
-
{MT:14.}
-
-
{MT:14.}
{MT:5.18}
-
-
-
-
-
MT-29

Advertisement