FujiFilm PROFECT CS Service Manual page 191

Table of Contents

Advertisement

Error
Error Name
Code
12225
Reading status error
12226
Operation line status error
12241
Scanner control board error (1)
12242
Scanner control board error (2)
12243
Scanner control board error (3)
12244
Scanner control board error (4)
Photomultiplier control board error
12251
(1)
Photomultiplier control board error
12252
(2)
12255
Leading-edge detection error (1)
Leading-edge detection timing
12256
error
12261
Polygon rotation error (1)
12262
Polygon rotation error (2)
011-202-11E
06.29.2012 FM6046
During routine processing, error is detected for the VSYNC signal on the SCN17A/SCT17B board.
During routine processing, error is detected for the SEDTM/FCLKTM signal on the SCN17A/SCT17B board.
During bootup or during routine processing, error is detected for the PLL function, polygon rotation function, start-
point interval function on the SCN17A/SCT17B board.
During routine processing, because the PMT high-voltage value is not set correctly on the SCN17A/SCT17B board,
error is detected for the high-voltage command diagnostics.
During bootup or during routine processing, error is detected for the start-point mask signal generated from the
SCN17A/SCT17B board and the index signal (PINDXL) from the polygonal mirror (POL).
During routine processing, an SCN board diagnostic error associated with polygon error is detected. Because the
PLL on the SCN17A/SCT17B board does not oscillate, a polygon sync signal cannot be detected.
During bootup or during routine processing, only error for the PMT analog power supply signal (+15VOKH,
+15VOKL) is detected (detail code: 008100). Note that if the HV is OFF, the detail code is 008000.
During bootup or during routine processing, only error for the high-voltage power supply signal (HVOKH, HVOKL) is
detected.
During bootup or during routine processing, only error for the leading-edge detection signal (SED1, SED2L) is
detected.
During routine processing, IP conveyance overrun occurs due to mechanical error or SCN17A/SCT17B board
control error, and only error for the SEDTM and FCLKTM signals on the SCN17A/SCT17B board is detected.
During bootup, error for the polygon lock signal (PONL, POKL) is detected so that the polygonal mirror function is
disabled. The start-point detection (SYN) and leading-edge detection (SED) functions are also disabled.
During bootup, error for the polygon lock signal (PONL, POKL) and error for the start-point interval function and
polygon rotation detection function on the SCN17A/SCT17B board are detected (detail code: 003400).
Occurrence Condition
CR-IR 363 Service Manual
MT-34
Analysis
Flow
{MT:18.}
{MT:18.}
{MT:5.19}
{MT:5.19}
{MT:5.19}
{MT:5.8}
{MT:5.12}
{MT:5.12}
{MT:5.14}
{MT:5.14}
{MT:5.8}
{MT:5.8}
MT-34

Advertisement

Table of Contents
loading

This manual is also suitable for:

Clearview csClearview csmCr-ir 363

Table of Contents