Make sure that "RESULT-OK" appears on screen.
6.
Using the X-ray dosimeter, measure the X-ray dose for exposing the
IP, and write it down.
7.
Expose with the tungsten X-ray tube (approx. 1 mR) for exposure even
if the IP is either of type ST (one side) or type ST-BD (both side).
8.
Read the IP exposed in step 7 for a one-minute period from 10 to 11
minutes after the exposure by the use of "TEST", "image format" or
"image format-2" and "SINGLE" in the CR console menu, and output
the image.
Make sure that the S value is normal with the measured X-ray dose and that
there is no particular irregularity.
REFERENCE
For 0.1 mR:
The S value is 2000.
For 1 mR:
The S value is 200.
For 10 mR:
The S value is 20.
INSTRUCTION
u
u
Be sure to follow the instruction mentioned in "20.2 Image/Conveyance Checks"
when outputting the image.
{MC:20.2_Image/Conveyance Checks}
9.
Make sure that "RESULT-OK" appears on screen.
011-202-11E
06.29.2012 FM6046
10.
Back up the data so adjusted into a floppy diskette.
{MC:19.1_Backing Up the Scanner Data}
CR-IR 363 Service Manual
MC-418
MC-418