FujiFilm PROFECT CS Service Manual page 274

Table of Contents

Advertisement

5.12
Error Code Analysis Flow: Scanner 12
12211, 12251, 12252, 13285, 13286
n Analysis Flow
m
Check the error log in the ERROR DB.
Q1
Errors 10230-10236 logged?
Y
Check the error code table again and troubleshoot.
N
{MT:2._Error Code Table}
A
Q2
Check the voltage of the SCN17A/SCT17B board test pin (CN5).
{MT:8._Procedures for Checking the Voltage}
Q3
IMG17A/B, CPU91A, and SCN17A/SCT17B boards tested in MUTL normal?
{MT:18._Board Test in MUTL}
N
Q6
Y
Q4
Check the photomultipliers (PMT).
{MT:13._Procedures for Checking the Scanner I/O}
Q5
Error recurs?
N
A
Y
m
Reseat the boards in the order named, and if the error is detected, replace the board.
1. SCN17A/SCT17B board
{MC:13.4_Scanner Board}
2. PMT17A/PMR17B board
{MC:11.3_PMT17A/PMT23A/PMT26A/PMT27B Board}
{MC:11.4_PMR17B/PMR23B/PMR26B/PMR27B Board}
3. IMG17A/B board
{MC:13.3_IMG17A/B Board}
4. CPU91A board
{MC:13.2_CPU91A Board}
011-202-11E
06.29.2012 FM6046
Q6
Reseat the error-causing board and check if the error occurs again.
N
A
Y
m
Replace the error-causing board.
{MC:13.2_CPU91A Board}
{MC:13.3_IMG17A/B Board}
{MC:13.4_Scanner Board}