FujiFilm PROFECT CS Service Manual page 180

Table of Contents

Advertisement

Error
Error Name
Code
Initialization feed path IP
10443
conveyance error (1) (UP)
Initialization feed path IP
10444
conveyance error (2) (DOWN)
Initialization feed entrance IP
10445
conveyance error (1) (DOWN)
Initialization feed entrance IP
10446
conveyance error (2) (DOWN)
Initialization side-positioning unit
10447
IP conveyance error (1) (DOWN)
Initialization side-positioning unit
10448
IP conveyance error (2) (UP)
Initialization side-positioning unit
10449
IP conveyance error (3) (UP)
011-202-04
11.15.2004 FM4425
During bootup, the IP transport motor (MB1, MB2) is driven to perform the remaining IP search processing; however,
the pre-BCR IP sensor (SB5) does not transition to OPEN within the specified period of time.
During bootup, the IP transport motor (MB1, MB2) is driven to perform the remaining IP search processing; however,
the pre-BCR IP sensor (SB5) does not transition to CLOSE within the specified period of time.
During bootup, the IP transport motor (MB1, MB2, MC3) is driven to perform the remaining IP search processing;
however, the IP feed/load sensor (SB1/2/3/4) does not transition to OPEN within the specified period of time.
During bootup, the IP transport motor (MB1, MB2, ME1) is driven to perform the remaining IP search processing;
however, the IP feed/load sensor (SB1/2/3/4) does not transition to OPEN within the specified period of time. (When
the IP in the feed/load conveyor unit is moved into the side-positioning conveyor unit.)
During bootup, the IP transport motor (MB1, MB2, MC3) is driven in the direction of subscanning conveyance to
perform the remaining IP search processing; however, the side-positioning entrance IP sensor (SG1) does not
transition to OPEN within the specified period of time.
During bootup, the IP transport motor (MC3) is driven in the direction of feed/load conveyor unit to perform the
remaining IP search processing; however, the side-positioning entrance IP sensor (SG1) does not transition to
CLOSE within the specified period of time.
During bootup, the IP transport motor (MC3) and subscanning motor (MZ1) are driven to perform the remaining IP
search processing; however, the side-positioning entrance IP sensor (SG1) does not transition to CLOSE within the
specified period of time.
Occurrence Condition
CR-IR 363 Service Manual
MT-23
Analysis
Flow
{MT:4.18}
{MT:4.18}
{MT:4.18}
{MT:4.18}
{MT:4.18}
{MT:4.19}
{MT:4.19}
MT-23