FujiFilm PROFECT CS Service Manual page 210

Table of Contents

Advertisement

Error
Error Name
Code
Driven-shaft grip release velocity
13413
correction process operation
13289
Format adjustment data error
Detection of communication
13520
abortion to CL
CL communication failure
13521
detected
13530
CL image data receive timeout
13629
Back-side HVCNT value error
Back-side start-point mask signal
13640
warning
Back-side main scan interval
13660
warning
Back-side HV command value
13685
error warning
Back-side HV detection value
13686
error warning
011-202-11E
06.29.2012 FM6046
When the grip drive motor (MZ2) was used in the routine processing or M-Utility mode to monitor the speed of a
driven shaft grip release operation other than for secondary erasure, the speed deviated from normal but remained
within the acceptable range. Therefore, the speed correction process was performed after termination of reading.
In the Maintenance Utility, when format adjustments are performed, out-of-range data is inputted.
During bootup or during routine processing, it is detected that the communication between the CL and the RU is
broken.
During bootup or during routine processing, because the communication between the CL and RU is broken, the
connection is suspended and initialization of communication between the CL and RU is performed again.
The CL was too late receiving the image data sent from the RU, so that the image was retransmitted from the RU.
The network is overloaded, or a communication error occurs due to hardware failure of the network, thereby causing
frequent retries.
During M-Utility operation, an out-of-spec HV is inputted.
- During bootup and routine processing, the start-point mask signal issued from the SCT12B board varies by 0.02%,
peak to peak.
- During bootup and routine processing, the detection of 12241 error is done with 2% variations.
During bootup or during routine processing, the scope of main-scan interval variations is outside the upper or lower
limit of the error determination value parameter.
<User operation>
When vibration is applied to the machine from the outside while the adjuster bolt is not lowered, the vibration is
transmitted to the scanning optics unit and logged.
<Action>
Check to ensure that the adjuster bolt is lowered.
During routine processing, error is detected for the HV command value signal sent from the SCT17B board to the
PMR17B board. The difference between MIN and MAX of the monitor value relative to the setup value is out of
specification.
During routine processing, error is detected for the HV detection value signal of the PMR17B board. The difference
between MIN and MAX of the monitor value relative to the setup value is out of specification.
Occurrence Condition
CR-IR 363 Service Manual
MT-53
Analysis
Flow
{MT:4.15}
-