10.5.1
Check Procedures via Monitoring Function (2)
1.
Start the RU PC-TOOL.
{MU:4.2.1_Starting RU PC-TOOL}
2.
Start monitoring of the RU PC-TOOL.
{MT:10.2.1_Starting the Monitoring Function}
3.
Navigate on the LCD panel of the RU to enter the "Erasure" Mode.
{Instruction Manual}
4.
Set a cassette corresponding to the sensor to be checked.
Cassette insertion
I/O No
destination
SA5
Shelf 1
SB1
SA11
Shelf 2
SB2
SA17
Shelf 3
SB3
SA23
Shelf 4
SB4
REFERENCE
To check the SG2 (11" x 14" size determination sensor), set a 11" x 14" size
cassette.
011-202-04
11.15.2004 FM4425
Cassette insertion
I/O No
destination
SB5
SB6
SB10
SC3
Set the cassette on any
SD2
shelf.
SE1
SE2
SG1
SG2
5.
Check the monitoring display.
Check the display of the monitoring window of the CL.
l Example of GOOD indication
l NO GOOD indication
For the sensor to be checked, neither "OPEN" nor "CLOSE" is displayed.
u NOTE u
Click on the [START] button in the monitoring window to initiate the monitoring
beforehand. Unless the monitoring is initiated, nothing is displayed in the window
even when the sensor changes its status.
6.
Exit the monitoring of the RU PC-TOOL.
{MT:10.2.2_Exiting the Monitoring Function}
CR-IR 363 Service Manual
MT-156
MT-156